dc.description.abstractThis thesis proposes an FMCW radar measurement algorithm for the extraction of time-varying respiratory data and heart rates based on phase measurement and a digital wavelet transform. In order to address the issue of sampling pauses resulting from hardware limitations, a piecewise constant interpolation of the inter-frame processing time was employed. Subsequently, a phase unwrapping algorithm was implemented and improved with regard to noise stability. To extract time-varying vital signs, the digital wavelet transform was employed in conjunction with a peak finding algorithm. The methodology was evaluated in a study involving three subjects and nine measurements under diverse conditions, including held breath and high-intensity measurements. The results demonstrated 100% precision for respiration rate and 96.5% precision for heart rate in relation to the reference measurement.en
dc.description.abstractIn dieser Arbeit wird ein FMCW-Radarmessalgorithmus für die Extraktion von zeitlich veränderlichen Vitalzeichen auf der Grundlage der Phasenmessung und der digitalen Wavelet-Transformation entwickelt. Um das Problem der durch Hardwarebeschränkungen bedingten Abtastpausen zu lösen, wird eine stückweise konstante Interpolation der Verarbeitungszeit zwischen den Frames verwendet. Anschließend wird ein Phase-Unwrapping-Algorithmus implementiert und im Hinblick auf die Rauschstabilität verbessert. Zur Extraktion der zeitvarianten Vitalparameter wird die digitale Wavelet-Transformation in Verbindung mit einem Peak-Finding-Algorithmus eingesetzt. Die Methode wurde in einer Studie mit drei Probanden und neun Messungen unter verschiedenen Bedingungen, einschließlich angehaltenem Atem und hochintensive Belastung, bewertet. Die Ergebnisse zeigten eine 100% Präzision für die Atemfrequenz und eine 96,5% Präzision für die Herzfrequenz in Bezug auf die Referenzmessung.de
dc.language.isoenen_US
dc.subjectkontaktlosen_US
dc.subjectVitalzeichenen_US
dc.subjectFMCWen_US
dc.subjectRadaren_US
dc.subjectSISOen_US
dc.subjectPhaseen_US
dc.subjectPhasenmessungen_US
dc.subject.ddc004: Informatiken_US
dc.subject.ddc600: Techniken_US
dc.subject.ddc610: Medizinen_US
dc.subject.ddc620: Ingenieurwissenschaftenen_US
dc.titleContactless Vital Sign Measurementen
dc.typeThesisen_US
openaire.rightsinfo:eu-repo/semantics/openAccessen_US
thesis.grantor.departmentDepartment Informations- und Elektrotechnik (ehemalig, aufgelöst 10.2025)en_US
thesis.grantor.universityOrInstitutionHochschule für Angewandte Wissenschaften Hamburgen_US
tuhh.contributor.refereeLeutelt, Lutz-
tuhh.identifier.urnurn:nbn:de:gbv:18302-reposit-222239-
tuhh.oai.showtrueen_US
tuhh.publication.instituteDepartment Informations- und Elektrotechnik (ehemalig, aufgelöst 10.2025)en_US
tuhh.publication.instituteFakultät Technik und Informatik (ehemalig, aufgelöst 10.2025)en_US
tuhh.type.opusBachelor Thesis-
dc.type.casraiSupervised Student Publication-
dc.type.dinibachelorThesis-
dc.type.driverbachelorThesis-
dc.type.statusinfo:eu-repo/semantics/publishedVersionen_US
dc.type.thesisbachelorThesisen_US
dcterms.DCMITypeText-
tuhh.dnb.statusdomainen_US
item.cerifentitytypePublications-
item.languageiso639-1en-
item.creatorOrcidKruse, Leevie Alexander-
item.advisorGNDKronauge, Matthias-
item.openairetypeThesis-
item.fulltextWith Fulltext-
item.openairecristypehttp://purl.org/coar/resource_type/c_46ec-
item.creatorGNDKruse, Leevie Alexander-
item.grantfulltextopen-
Enthalten in den Sammlungen:Theses
Dateien zu dieser Ressource:
Datei Beschreibung GrößeFormat
BA_Contactless Vital Sign Measurement_geschwärzt.pdf1.29 MBAdobe PDFÖffnen/Anzeigen
Zur Kurzanzeige

Seitenansichten

131
checked on 30.12.2025

Download(s)

31
checked on 30.12.2025

Google ScholarTM

Prüfe

HAW Katalog

Prüfe

Feedback zu diesem Datensatz


Alle Ressourcen in diesem Repository sind urheberrechtlich geschützt.