| DC Field | Value | Language |
|---|---|---|
| dc.contributor.advisor | Kronauge, Matthias | - |
| dc.contributor.author | Kruse, Leevie Alexander | - |
| dc.date.accessioned | 2025-11-05T08:28:39Z | - |
| dc.date.available | 2025-11-05T08:28:39Z | - |
| dc.date.created | 2024-09-10 | - |
| dc.date.issued | 2025-11-05 | - |
| dc.identifier.uri | https://hdl.handle.net/20.500.12738/18350 | - |
| dc.description.abstract | This thesis proposes an FMCW radar measurement algorithm for the extraction of time-varying respiratory data and heart rates based on phase measurement and a digital wavelet transform. In order to address the issue of sampling pauses resulting from hardware limitations, a piecewise constant interpolation of the inter-frame processing time was employed. Subsequently, a phase unwrapping algorithm was implemented and improved with regard to noise stability. To extract time-varying vital signs, the digital wavelet transform was employed in conjunction with a peak finding algorithm. The methodology was evaluated in a study involving three subjects and nine measurements under diverse conditions, including held breath and high-intensity measurements. The results demonstrated 100% precision for respiration rate and 96.5% precision for heart rate in relation to the reference measurement. | en |
| dc.description.abstract | In dieser Arbeit wird ein FMCW-Radarmessalgorithmus für die Extraktion von zeitlich veränderlichen Vitalzeichen auf der Grundlage der Phasenmessung und der digitalen Wavelet-Transformation entwickelt. Um das Problem der durch Hardwarebeschränkungen bedingten Abtastpausen zu lösen, wird eine stückweise konstante Interpolation der Verarbeitungszeit zwischen den Frames verwendet. Anschließend wird ein Phase-Unwrapping-Algorithmus implementiert und im Hinblick auf die Rauschstabilität verbessert. Zur Extraktion der zeitvarianten Vitalparameter wird die digitale Wavelet-Transformation in Verbindung mit einem Peak-Finding-Algorithmus eingesetzt. Die Methode wurde in einer Studie mit drei Probanden und neun Messungen unter verschiedenen Bedingungen, einschließlich angehaltenem Atem und hochintensive Belastung, bewertet. Die Ergebnisse zeigten eine 100% Präzision für die Atemfrequenz und eine 96,5% Präzision für die Herzfrequenz in Bezug auf die Referenzmessung. | de |
| dc.language.iso | en | en_US |
| dc.subject | kontaktlos | en_US |
| dc.subject | Vitalzeichen | en_US |
| dc.subject | FMCW | en_US |
| dc.subject | Radar | en_US |
| dc.subject | SISO | en_US |
| dc.subject | Phase | en_US |
| dc.subject | Phasenmessung | en_US |
| dc.subject.ddc | 004: Informatik | en_US |
| dc.subject.ddc | 600: Technik | en_US |
| dc.subject.ddc | 610: Medizin | en_US |
| dc.subject.ddc | 620: Ingenieurwissenschaften | en_US |
| dc.title | Contactless Vital Sign Measurement | en |
| dc.type | Thesis | en_US |
| openaire.rights | info:eu-repo/semantics/openAccess | en_US |
| thesis.grantor.department | Department Informations- und Elektrotechnik (ehemalig, aufgelöst 10.2025) | en_US |
| thesis.grantor.universityOrInstitution | Hochschule für Angewandte Wissenschaften Hamburg | en_US |
| tuhh.contributor.referee | Leutelt, Lutz | - |
| tuhh.identifier.urn | urn:nbn:de:gbv:18302-reposit-222239 | - |
| tuhh.oai.show | true | en_US |
| tuhh.publication.institute | Department Informations- und Elektrotechnik (ehemalig, aufgelöst 10.2025) | en_US |
| tuhh.publication.institute | Fakultät Technik und Informatik (ehemalig, aufgelöst 10.2025) | en_US |
| tuhh.type.opus | Bachelor Thesis | - |
| dc.type.casrai | Supervised Student Publication | - |
| dc.type.dini | bachelorThesis | - |
| dc.type.driver | bachelorThesis | - |
| dc.type.status | info:eu-repo/semantics/publishedVersion | en_US |
| dc.type.thesis | bachelorThesis | en_US |
| dcterms.DCMIType | Text | - |
| tuhh.dnb.status | domain | en_US |
| item.advisorGND | Kronauge, Matthias | - |
| item.languageiso639-1 | en | - |
| item.openairetype | Thesis | - |
| item.fulltext | With Fulltext | - |
| item.openairecristype | http://purl.org/coar/resource_type/c_46ec | - |
| item.creatorOrcid | Kruse, Leevie Alexander | - |
| item.cerifentitytype | Publications | - |
| item.creatorGND | Kruse, Leevie Alexander | - |
| item.grantfulltext | open | - |
| Appears in Collections: | Theses | |
Files in This Item:
| File | Description | Size | Format | |
|---|---|---|---|---|
| BA_Contactless Vital Sign Measurement_geschwärzt.pdf | 1.29 MB | Adobe PDF | View/Open |
Note about this record
Export
Items in REPOSIT are protected by copyright, with all rights reserved, unless otherwise indicated.